PRINCIPLE OF OPERATION: Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
-
INITIALIZATION TIME: Capnogram within 15 sec., full specifications within 60 sec.
-
CALIBRATION: Routine calibration is not required. 15 sec. adapter zero performed when changing to different style of adapter.
-
ACCURACY VERIFICATION: Accuracy verifier provided on sensor cable.
-
CAPNOSTAT MAINSTREAM CO2 SENSOR:
Size: 1.30 × 1.67 × 0.85 inches (3.30 × 4.24 × 2.16 cm);Weight: 0.64 oz. (18 grams) - cable excluded
Shock Resistance: Withstands repeated 6 foot drops.
-
AIRWAY ADAPTERS: Adult/pediatric reusable or single patient use (added dead space <5cc), neonatal reusable (added dead space <.5cc), and sampling adapter for non-intubated patients.
-
SAMPLING FLOW RATE : 180 ml/min (for non-intubated patients).
-
CO2 (CARBON DIOXIDE):
Compensations: N2O & O2 (selectable), BP (automatic)
Capnogram: Selectable sweep speed (12.5 or 25 mm/sec) Range: 0 - 100 mmHg.
Accuracy: ±2 mmHg (for 0 - 40 mmHg) ±5% of reading (for 41 - 70 mmHg) ±8% of reading (for 71 - 100 mmHg)
Resolution: 1 mmHg Response Time: 60 ms
-
RESPIRATORY RATE:
Accuracy: ±1 br/min
Range: 0-150 br/min
Resolution: 1 br/min
-
ALERTS:
Limits: Adjustable limits for ETCO2, SpO2, Resp. and Pulse Rates
Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
Visual: On-screen & red "Alert Bar"
-
DISPLAY
Type: Dot matrix, Cold Cathode Display (CCD)
Size: 5" W × 1.5" H (12.7 × 3.08 cm)
-
GRAPHIC TREND/HISTOGRAM
Memory: 24 hr., battery backed
Format: On-screen or printed, in 30 min., 2, 8, or 12 hr. segments
-
COMMUNICATION OUTPUT
Digital: RS232 (printer, NovaCARD, computer interface)
Analog: Optional module, 0-1VDC (numerics & waveforms)
-
PHYSICAL
Size: 3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm)
Weight: 8 lb (3.63 kg)
-
ELECTRICAL
Power requirements: 100-120/200-240 VAC, 50-60 Hz, 30 VA
Battery: Sealed lead acid gel cell, 2 hr. life, 12 hr. recharge
-
STAND-ALONE CAPNOGRAPH CONFIGURATION: The CAPNOGARD capnograph has the same technical specifications as the CO2SMO, excluding the pulse oximeter